Los hornos de mufla LT 3/11 - LT 60/11 son la opción ideal para el uso diario en el laboratorio. Estos modelos de alta calidad tienen un diseño atractivo y avanzado, y son muy fiables. Disponen de una puerta levadiza estándar.
- Carcasa de acero inoxidable texturada.
- Carcasa doble para temperaturas externas extremas y alta estabilidad
- Puerta levadiza (modelos L) que mantiene la superficie caliente alejada del operador
- Salida de aire en la pared posterior del horno
- Relés de estado sólido que ofrecen un funcionamiento con bajo ruido
- Aplicación definida dentro de las limitaciones de las instrucciones de uso
- NTLog Basic para el controlador Nabertherm: registro de datos de proceso con unidad flash USB
Heating is from two sides by means of ceramic heating plates (heating from three sides for muffle furnaces LT 24/11 - LT 60/11). These ceramic heating plates with an integral heating element are protected from fumes and splashing, and are easy to replace. Only fibre materials which are not classified as carcinogenic according to TRGS 905, class 1 or 2 are used.
The furnace is equipped with a B 510 standard controller with touch screen display, featuring five freely adjustable programs, four segments and two extra functions. These controllers are PID microprocessor controlled with a self-diagnostic system and integrated process documentation. The B 510 controllers can automatically activate the vent with fan or catalyst (not model LT 15).
Información para pedidos: All units require a 230 V single-phase power supply, except LT 24, LT 40 and LT 60 (400 V 3/N/PE). Optional accessories e.g. a vent, vent with fan or catalytic converter, over-temperature limit controller with adjustable temperature cut out for thermal protection Class 2, in accordance with EN 60519-2 as temperature limiter to protect the oven and load, protective gas connection on the rear wall of furnace. Manual or automatic gas supply systems are also available. Corresponding models with fold down doors, different Tmax and other controllers are also available. Please search the website for more details.
Precaución: Maximum continuous operating temperature should be 50 °C below the 'Max. temp.' of the model.
Note heat up times: If connected at 230 V 1/N/PE rsp. 400 V 3/N/PE.
Outer dimensions include opened lift-up door.
* Temperature uniformity of ±5 K in the empty work space for temperatures >800 °C